Arrester NZR-7B
An arrester protects measuring instruments and transducers from induced lighting. In lighting-prone areas, the induced voltage of lighting traveling through the cables of measuring instruments or switch boxes can damage instruments and transducers. When placing measuring instruments and transducers, connection of NZR-7B/NZ-7C on cabling between an instrument and a switch box or NZ-6B on cabling between a transducer and switch box will protect the instrument and transducer by allowing induced current to pass to the ground.
end of production
- Features
- While a measuring instrument is on standaby, NZR-7B automatically separates the circuit between all instruments with a high-voltage relay to block induced current
- Either soldering or NDIS plug is possible for cable connection
-
Specifications
Type NZR-7B Applicable instruments Data logger: TDS-630, TDS-540, TDS-530
Switching Box: ASW-50C, ASW-30C, SSW-50DSurge resistance (8/20µs)
Between bridges: 442A between bridge
Between bridge and ground: 10kA
Between F and ground: 10kA
Between G and ground: 10kADimensions 290 (W) x 64 (H) x 101 (D) mm Weight Approx. 1.4kg