AWMD-5, AWMD-8 (for dynamic measurement only : -196 to +800℃) Full bridgeCE_logo

The AWMD is applicable up to 800°C and is dedicated to dynamic measurement. A high pass filter is a standard accessory. Using the high pass filter, unnecessary direct current component and low frequency component (thermal output, drift, etc.) are cancelled. A DC exciting dynamic strainmeter such as DC-96A/DC-97A, smart strain recorder such as DC-204R, or multi-recorder TMR-200 series should be used for the measurement.

Main object material Metal
Operational temperature -196 to +800℃
Temperature compensation range -
Bonding adhesive Spot welding (TML spot welder W-50R is required)
Backing Inconel 600
Element Special alloy
Strain limit 1% (10000 x 10-6 strain)
Fatigue life at room temperature 1 to 106 (±1000 x 10-6 strain)
  • Specifications

    awmd-5-8
    Type Gauge
    length
    (mm)
    Gauge base Test specimen Resistance in (Ω)
    Dimensions (mm) Material
    AWMD-5-AKM-2(6F)-1.6Hz 5 L10 x W3 x T0.7 Inconel 600 Inconel 600 equivalent 60
    AWMD-8-A-2-1.6Hz 8 L16 x  W5 x T0.7 120

    Each package contains one strain gauge.
    Other cable length available on request 
    (Made to order by every 1 m for MI cable and by every 0.5 m for vinyl cable)