Arrester NZ-7C

An arrester protects measuring instruments and transducers from induced lighting. In lighting-prone areas, the induced voltage of lighting traveling through the cables of measuring instruments or switch boxes can damage instruments and transducers. When placing measuring instruments and transducers, connection of NZR-7B/NZ-7C on cabling between an instrument and a switch box or NZ-6B on cabling between a transducer and switch box will protect the instrument and transducer by allowing induced current to pass to the ground.

Arrester NZ-7C
Features
  • Either soldering or NDIS plug is possible for cable connection
  • Contains a commercial power supply (built-in varistor) terminal, allowing power supply for a switch box to be connected easily
  • Specifications

    Type NZ-7C
    Applicable instruments Data logger: TDS-630, TDS-540, TDS-530
    Switching Box: ASW-50C, ASW-30C, SSW-50D
    Surge resistance (8/20µs)
    Between bridges: 442A
    Between bridge and ground: 10kA
    Between F and ground: 442A
    Between G and ground: 442A
    Dimensions 290 (W) x 62(H) x 101 (D) mm
    Weight Approx. 1.3kg