Network Module
Network Module
Strain Quarter Bridge Module NSW-011C
This is a low consumption module dedicated to quarter bridge 3-wire connection. The extension distance and number of connection are improved compared to conventional modules.
-
Specifications
Type NSW-011C Number of measurement point 1 point Measurement time 200 ms/points (including transfer time) Programmable item Channel number (000 to 999) Measurement mode Strain quarter bridge 3-wire method Applicable gauge resistance Either 120 Ω or 350 Ω (to be specified when ordering) Measuring range ±30000 x 10-6 strain
Dimensions 50(W) x 20(H) x 100(D) mm Weight Approx. 250 g
Network Module
Strain Full Bridge Module NSW-014C
This is a low comsumption module dedicated to full bridge connection including strain gauge based transducer. The extension distance and number of connection are improved compared to conventional modules.
-
Specifications
Type NSW-014C Number of measurement point 1 point Measurement time 200 ms/point (including transfer time) Programmable item Channel number (000 to 999) Measurement mode Strain full bridge Applied gauge resistance 120 to 1000 Ω Measuring range ±30000 x 10-6 strain
Dimensions 50(W) x 20(H) x 100(D) mm Weight Approx. 250 g
Network Module
Strain Full Bridge 2-Channel Module NSW-024C
This is a low-consumption, 2-channel network module for full bridge strain gauge-type transducers.
-
Specifications
Type NSW-024C Number of measurement point 2 points Measurement time 200 ms/point (including transfer time) Programmable item Channel number for input 0 : 000 to 998
Channel number for input 1 : INPUT +1Measurement mode Strain full bridge Applied gauge resistance 120 to 1000Ω Measuring range ±30000 x 10-6 strain Dimensions 95(W) x 20(H) x 100(D) mm Weight Approx. 300 g
Network Module
Counter Module NSW-01CC
A module that counts precipitation, flow rate, quantity, number of vehicles that pass by, frequency of machine operation, and other information with no-voltage contact or open collector input.
A built-in backup battery allows to continue counting even when the power supply is not available.
-
Specifications
Type NSW-01CC Measured object Precipitation meter, etc. Number of measurement point 1 point Measurement time 200 ms/point (including transfer time) Programmable item Channel number (000-999) Input signal Non-voltage contact / Open collector signal Square Wave Responding pulse width 0.01 sec or more Measuring range 0 ~ 31999 counts Measurement data Accumulated counts Reset At power on / Full count / External rese Back up Alkaline AA dry batteries x 3 pcs. Approx. 3 months Dimensions 95(W)×35(H)×100(D)mm Weight Approx. 250 g (Alkaline AA dry batteries x 3 pcs. 70 g not included) Note: The counter module NSW-01CC is not a low-consumption type, but four units of NSW-01CC can be connected when 100 units are connected, and the total length is 2,000 m.
*Connection of four units and more is possible depending on the conditions.