Data Logger TS-560
T-ZACCS5 Data Logger TS-560
Next-generation data logger with sophisticated functions integrated in a compact body
T-ZACCS 5 TS-560 is a data logger that is easily capable of multi-point measurement when combined with high speed switching box IHW-50G and/or switching box ISW-50G. The maximum number of measuring point is 1000, and an arbitrary combination of inputs including strain gauges, strain gauge type transducers, dc voltage, thermocouples and Pt-RTDs are available according to the measurement conditions. Since it has achieved smaller and more lightweight construction than conventional data loggers, it is applicable to various measurement situations. The color LCD touch screen provides comfortable and intuitive operation and clear monitoring display. In addition to real-time monitoring, it is possible to make quick confirmation/verification of measured data after the measurement.
- Features
- Data are securely held by the robust built-in data memory
- Stable interval measurement possible at clock accuracy of daily rate ±1 second
- Batch setting of coefficient, unit, decimal point and sensor type using “Sensor ID”
- “Quick interval” function provided for interval measurement without the need for setting
- “Remote data logger” function provided for control through an internet browser
- Number of measuring point is 1000 at maximum (using external switching boxes)
- Equipped with Complete Compensation Method of Strain
- Various check functions (automatic checking using timer is possible)
- Touch screen color LCD monitor
- Automatic measurement function using interval timer and monitor comparator
- SD card and USB memory are accepted as recording media
- Equipped with LAN, USB and RS-232C as standard specifications
- Powered by built-in battery even if external power supply is interrupted
-
Specifications
Type TS-560 Number of measuring point 1000 points at maximum Measuring speed IHW-50G 1 s/1000 points ISW-50G 3 s/1000 points Measurement mode Initial, Direct, Measure (Direct only for temperature measurement) Simple measure Coefficient 1.000 Unit Depends on sensor mode Decimal point Depends on sensor mode Compensation mode Comet NON/Comet A/Comet B Sensor mode Strain Quarter bridge 3-wire 120/240/350 Ω
Half bridge, Half bridge common dummy
Full bridge
Full bridge high resolution Full bridge constant current 350 Ω
Full bridge constant current high resolution 350 Ω
Full bridge 0-2V mode
Temperature-integrated strain gauge 120/240/350 Ω
1-gauge 4-wire 120/240/350 ΩTemperature Thermocouple T, K, J, B, S, R, E, N
Pt-RTD Pt100 3WVoltage 640 mV, 64 V Recording Internal memory, SD card, USB memory Interface LAN, USB, RS-232C External dimensions 200(W) × 30(H) × 150(D) mm Weight Approx. 1 kg Power supply AC adaptor (CR-1867) AC 100~240 V 50/60 Hz 152 VA max. Built-in battery Built-in battery Lithium-ion battery Continuous operating time Approx. 1 hour (Temperature: 23°C±5°C, When IHW-50G is connected)